Specifications

Thermocouple Reference Junction Adjustments
These adjustments are plug-in module specific and only affect thermo-
couple measurements. The calibration constants created by these
adjustments are stored in non-volatile memory on the plug-in module.
1 Connect a 10 k (YSI 44031) thermistor to each of the following channels
(a kit of five thermistors is available as Keysight part number 34308A):
For the 34901AChannels 6 and 17
For the 34902AChannels 6 and 11
For the 34908AChannels 6 and 16
Keep the thermistor leads as short as possible. Locate the thermistor
as near to the input connectors as possible.
2 Install the plug-in module in the mainframe in slot 200. Apply power
and allow a 2 hour warm-up.
3 Set 10 k thermistor measurements on Channels 206 and 217
(or 206 and 211). Before executing each test, you must press
to enable reading monitoring on the selected channel (or use the
ROUTe:MON command from the remote interface).
4
Press to enter the calibration menu. Press again to
begin the adjustment procedure.
5 Verify the adjustment (see page 112).
Module Reference
34901A
4
Chapter 4 Calibration Procedures
Thermocouple Reference Junction (Optional)
113