Specifications
104
Chapter 4 Calibration Procedures
Relay Verification
...Continued from previous page
Tests 9 - 16
(ROW2):
Move the HI, LO, Sense HI and Sense LO leads to ROW2 as shown
below.
Test Lead Connections for ROW2
Press and hold down until “Card Reset” is displayed on the front
panel.
For each test, close only the channels shown in the “Channels Closed”
column below. Record the 4-wire ohms measurements from the external
DMM in the following table.
*Only the channel currently under test should be closed at one time. All other channels should be open.
Continued on next page...
Test
#
Channel
Closed*
Measured Value
Test
Limit
Relay
Measured
9 Ch 21 & 41 ________Ohms 2.00Ω K201
10 Ch 22 & 42 ________Ohms 2.00Ω K202
11 Ch 23 & 43 ________Ohms 2.00Ω K203
12 Ch 24 & 44 ________Ohms 2.00Ω K204
13 Ch 25 & 45 ________Ohms 2.00Ω K205
14 Ch 26 & 46 ________Ohms 2.00Ω K206
15 Ch 27 & 47 ________Ohms 2.00Ω K207
16 Ch 28 & 48 ________Ohms 2.00Ω K208