Specifications

103
Chapter 4 Calibration Procedures
Relay Verification
4
Module Reference
Tests 1-8 (ROW1): Open all channels on the module by performing a Factory Reset (press
and select “Recall State”; press
again and select “Factory
Reset”).
For each test, close only the channels shown in the “Channels Closed”
column below (module in slot 200). Record the 4-wire ohms
measurements from the external DMM in the following table.
*Only the channel currently under test should be closed at one time. All other channels should be open.
Continued on next page...
Test
#
Channel
Closed*
Measured Value
Test
Limit
Relay
Measured
1 Ch 11 & 41 ________Ohms 2.00Ω K101
2 Ch 12 & 42 ________Ohms 2.00Ω K102
3 Ch 13 & 43 ________Ohms 2.00Ω K103
4 Ch 14 & 44 ________Ohms 2.00Ω K104
5 Ch 15 & 45 ________Ohms 2.00Ω K105
6 Ch 16 & 46 ________Ohms 2.00Ω K106
7 Ch 17 & 47 ________Ohms 2.00Ω K107
8 Ch 18 & 48 ________Ohms 2.00Ω K108