User`s manual

80
Chapter 4 Features and Functions
Scanning
4
Configuring a Scan
You can use delay times to control a scan.
You can specify a delay time (from 0 to 99999.999 seconds, with 1 ms
resolution) between when a channel is closed and when the next
operation begins (and, if configured, a trigger out pulse is sent). The
delay time can be set individually for each channel in the scan list, or
one delay can be set for all channels in the scan list. Note that, even if
the trigger source is set to IMMediate, the scan list will not advance
until the delay time is met.
Front-Panel Operation: Use the S.List menu to set the scan delay
time as shown below.
Remote Interface Operation:
OUTPUT 709; “DELAY 2000; CHAN 101”
! Set a 2 second delay time
00000.0000 SECS
Use arrow keys and knob to
select the delay time. Press Enter.
DELAY TIME
Turn the knob to select DELAY TIME.
Press Enter.
3499
Press S.List to exit the menu.