Chapter 3 System Overview
3488 Mode Differences
4
3
3488 Mode Differences
This manual describes using the Agilent 3499A/B/C in the 3488 mode of
operation. In general, the SCPI mode of operation is recommended. All
the features and functions of 3499A/B/C system and plug-in modules are
supported in SCPI mode. The 3488 mode is a subset and does not support
all available features.
Some of the major differences in 3488 mode of operation are summarized
below:
• The RS-232 interface is not supported in 3488 mode. You must use
the GPIB interface for remote programming.
• The 3488 mode does not provide as much control over the scanning
process as SCPI mode. You cannot use the arm and trigger layers of
scan control.
• The 3488 mode only provides access to the status byte register. You
cannot use the Standard Event or Operation Status registers.
• With firmware revision 4.0, you can store up to 40 setups (50 setups
in SCPI mode).
• Scan lists are not saved after cycling power.
• The relay cycle count feature is only available from the front-panel.
• The N2260A and N2266A can only be used in 40-channel 2-wire
mode.
• The 3488 Mode does not support the following plug-in modules:
N2272A, N2276A/B, or N2282A.
• Port numbering for Digital I/O modules is different.