User`s manual
28
Chapter 2 Front-Panel Operation
To Monitor a Channel or a Slot
2
To Monitor a Channel or a Slot
You can continuously monitor the current status of a particular
switching channel, a digital I/O port, or an entire plug-in module.
Monitoring from the front panel is especially useful when developing and
debugging remote interface commands.
1. Press the monitor key, the MON annunciator lights up to indicate the
instrument is in the monitoring state.
2. Select the slot or the channel/port to be monitored. The displayed
information depends on the selected module type. Typical displays
are shown in the table on page 76.
3. If only part of the channel status on the module can be displayed at
one time, press Enter to display the next part.
For multiplexer modules and GP Relay modules, 10 channels can be
displayed at one time; for matrix modules, one Row or one Column
can be displayed at one time; for digital I/O modules, two 8-bit ports
can be displayed at one time. For multifunction modules, the first
function on the module is displayed, then the next.
4. Press the monitor key again to end monitoring (the MON
annunciator turns off).
Note The built-in digital I/O bits/port (on the controller module) can be
monitored either individually as bit channels (numbered 091 through
094) or as a 4-bit port (numbered 090). However, the individual bit
channels on a digital I/O or multifunction module (with a DIO function)
cannot be monitored.