Chapter 5 Remote Interface Reference
3488 Mode Commands
4
5
RECALL <1-40>
Recalls the channel/bit setup stored by the command STORE. Only relays
and static Digital Outputs are recalled. This means that only those
channels/bits that are closed in the stored channel setup will be closed
when the channel setup is recalled.
Note If no state has been stored in the channel setup selected, an error will
result and the 3499A/B/C will send SRQ if it was enabled (see MASK on
page 122). In this case, the 3499A/B/C remains in its previous state.
OUTPUT 709; “CLOSE 103,104,202,204,206; STORE 6”
OUTPUT 709; “RECALL 6”
RESET
Used to reset the 3499A/B/C to the power on state. Reset performs the
following:
• Identifies all plug-in modules and resets them (all relays open)
• Sets all parameters of the instrument to their default conditions. The
reset conditions are shown on page 71.
This command does not affect the GPIB settings or stored states.
Note If RESET is executed over the remote interface, interface functions are
not affected (i.e. the 3499A/B/C will stay in Remote mode). If the front
panel RESET key is pressed, interface functions are reset.
OUTPUT 709; “RESET”