User`s manual

106
Chapter 5 Remote Interface Reference
3488 Mode Commands
5
CMON <slot>
Causes the 3499A/B/C to monitor the individual card (module) in the
specified slot.
The slot parameter is a decimal ranging from 0 to 9 (mainframe
dependent). Valid slot numbers are:
Specifying 0 as the slot number turns off the monitor function. Cycling
power on the instrument or resetting the modules will cancel the
monitoring.
You cannot monitor the 4-bit built-in DIO, an individual DIO port, or a
switching channel.
The displayed information is module dependent, refer to page 75 for
more details.
OUTPUT 709; “CMON 2”
ENTER 709; A$
DISP A$
CRESET <slot>[,<slot>...]
Resets all channels/bits on the specified modules to their open state.
Only the modules in the specified slots are affected.
The slot parameter is a decimal ranging from 0 to 9 (mainframe
dependent). Valid slot numbers are:
Switching Modules: All the switching channels on the specific module
are reset to their open state.
DIO Modules: A digital I/O module is reset to its default mode. That
means the bits on the module are set to open (high impedance), and the
module is in the Static Mode #1 (no handshaking).
OUTPUT 709; “CRESET 2”
! Reset the module in slot 2.
3499A slots 0 through 5
3499B slots 0 through 2
3499C slots 0 through 9
3499A slots 0 through 5
3499B slots 0 through 2
3499C slots 0 through 9