User`s manual
104
Chapter 5 Remote Interface Reference
3488 Mode Commands
5
3488 Mode Commands
CHAN [<channel_address>]
Opens the last channel/bit closed by either STEP or CHAN command
and then closes the specified channel/bit.
The channel_address has the form @snn, where s is the slot number and
nn is the channel number. For all mainframes, slot 0 refers to the 3499A/
B/C control board. Valid slot numbers are:
The channel number, nn, is plug-in module dependent. Channel
numbers for plug-in modules are listed on page 70.
If no channel/bit is specified with CHAN, the 3499A/B/C will respond with
the number of the last channel closed. If no channel has been closed since
the last reset, then CHAN returns the number 0.
If a Scan List is in effect when CHAN is executed, the 3499A/B/C will
search through the list and position the Scan List pointer to the channel
specified by CHAN. If CHAN specifies a channel that is not in the Scan List,
and then STEP is executed, the 3499A/B/C will open the channel closed
by CHAN and go to the beginning of the Scan List.
OUTPUT 709; “CHAN 203”
! Close channel 203.
OUTPUT 709; “CHAN
ENTER 709; A
DISP A
3499A slots 0 through 5
3499B slots 0 through 2
3499C slots 0 through 9