Specifications
4-10 Chapter 4
Appendix
DANL Manual Performance Verification
-
Table 4-9 Agilent E4440A Performance Verification Test Record
Agilent Technologies
Model E4440A Report No. ___________
Serial No. ___________ Date ___________
Test Description Minimum
Results
Measured
Maximum
Measurement
Uncertainty
Displayed Average
Noise Level
Note: Enter results with preamp on in the appropriate section based upon
the ambient temperature when the test was performed.
Preamp On
100 kHz (1)________ –159 dBm ±0.07 dB
199 kHz (2)________ –159 dBm ±0.07 dB
201 kHz (3)________ –159 dBm ±0.07 dB
499 kHz (4)________ –159 dBm ±0.07 dB
501 kHz (5)________ –163 dBm ±0.07 dB
9.9 MHz (6)________ –166 dBm ±0.07 dB
10.1 MHz (7)________ –169 dBm ±0.07 dB
1.0 GHz (8)________ –168 dBm ±0.07 dB
1.2 GHz (9)________ –167 dBm ±0.07 dB
2.4 GHz (10)________ –165 dBm ±0.07 dB
2.6 GHz (11)________ –165 dBm ±0.07 dB
3.0 GHz (12)________ –165 dBm ±0.07 dB
Preamp Off
10 kHz (13)________ –134 dBm ±0.07 dB
99 kHz (14)________ –134 dBm ±0.07 dB
101 kHz (15)________ –144 dBm ±0.07 dB
990 kHz (16)________ –144 dBm ±0.07 dB
1.01 MHz (17)________ –149 dBm ±0.07 dB
500 MHz (18)________ –154 dBm ±0.07 dB
1.1 GHz (19)________ –154 dBm ±0.07 dB
1.3 GHz (20)________ –153 dBm ±0.07 dB
2.0 GHz (21)________ –153 dBm ±0.07 dB
2.4 GHz (22)________ –153 dBm ±0.07 dB