Specifications
4-4 Chapter 4
Appendix
DANL Manual Performance Verification
-
12.Press Single and wait for 100 averages.
13.Press Display, Display Line, On.
14.Scroll the display line so that it bisects the trace. Read the display line amplitude and
subtract 30 from the value. Record the result under “Normalized DANL” in the table.
15.Repeat step 12 through step 15 for all frequencies listed in Table 4-2.
16.Record the results of Table 4-2 in the test record table for your model instrument later
in this chapter. Refer to Table 4-7 on page 4-7. If the analyzer is an E4443A stop here.
17.Repeat step 12 through step 15 for all frequencies listed in Table 4-3.
18.Record the results of Table 4-3 in the test record table for your model instrument later
in this chapter. Refer to Table 4-7 on page 4-7. If the analyzer is an E4445A stop here.
1.1 GHz 19)
1.3 GHz 20)
2.0 GHz 21)
2.4 GHz 22)
2.6 GHz 23)
3.0 GHz 24)
3.1 GHz 25)
4.0 GHz 26)
5.0 GHz 27)
6.5 GHz 28)
6.7 GHz 29)
Table 4-3 DANL (PSA E4445A, E4440A, E4446A, E4448A)
Frequency Normalized DANL
8.0 GHz 30)
9.0 GHz 31)
10.0 GHz 32)
11.0 GHz 33)
12.0 GHz 34)
13.1 GHz 35)
Table 4-2 DANL (All PSA Instruments)
Frequency Normalized DANL