Specifications

Chapter 4 4-3
Appendix
DANL Manual Performance Verification
5. Press Single and wait for 100 averages.
6. Press Display, Display Line, On.
7. Scroll the display line so that it bisects the trace. Read the display line amplitude and
subtract 30 from the value. Record the result under “Normalized DANL” in the table.
8. Repeat step 5 through step 8 for all frequencies listed in Table 4-1.
9. Record the results of Table 4-1 in the test record table for your model instrument later
in this chapter. Refer to Table 4-7 on page 4-7.
10.Press [AMPLITUDE], More 1 of 3, Int Preamp, Off.
11.Press [FREQUENCY] and enter the first frequency listed in Table 4-2.
Table 4-1 DANL (PSA with Option 1DS)
Frequency Normalized
DANL
100 kHz 1)
199 kHz 2)
201 kHz 3)
499 kHz 4)
501 kHz 5)
9.9 MHz 6)
10.1 MHz 7)
1.0 GHz 8)
1.2 GHz 9)
2.4 GHz 10)
2.6 GHz 11)
3.0 GHz 12)
Table 4-2 DANL (All PSA Instruments)
Frequency Normalized DANL
10 kHz 13)
99 kHz 14)
101 kHz 15)
990 kHz 16)
1.01 MHz 17)
500 MHz 18)