Specifications

Chapter 3 3-9
User Interface
Flatness Compensation Table
Entering the Flatness Correction Data
To manually enter these recorded values into the PSA instrument, perform the following
sequence for each point.
[AMPLITUDE] > More 1 of 3 > Corrections > Other > Edit
Select [FREQUENCY] 2.849 GHz
Select [AMPLITUDE] 0 dB
After each entry, the Point counter automatically increments for entry of the next point.
Select [FREQUENCY] 2.85 GHz
Select [AMPLITUDE] ___ dB (Refer to Table 3-4)
Continue entering data for each point in the table.
Press [RETURN], when finished.
To use the new correction table, insure that the corrections are turned on.
CAUTION
It is recommended that the correction set entered be saved on the
internal drive or a floppy disk for future references. See the PSA User’s
and Programming Reference Guide for information on saving correction
values.
Table 3-4 A Sample Frequency Amplitude Sheet
Point Frequency
(GHz)
Delta
Amplitude (dB)
Correction
(dBm
1
)
1. The correction field will contain the negative, or opposite of the Ampli-
tude field. The correction value is what the instrument uses rather
than the Amplitude (or error) value.
1
2.849
2
2. This point is needed so that the corrections are not applied to Band 0.
0 dB 0 dB
2 2.85 ___ dB – (___ dB)
___ GHz ___ dB – (___ dB)
200
___ GHz
3
3. This should be the upper limit of the PSA.
___ dB – (___ dB)