Specifications

3-6 Chapter 3
User Interface
Flatness Compensation Table
Preamp Compensation Table Command Interface
The following command turns on or off the correction system.
[:SENSe]:CORRection:CSET:ALL[:STATe] OFF|ON|0|1
The following command turns on or off the correction file to be used
1
.
[:SENSe]:CORRection:CSET3:[:STATe] OFF|ON|0|1
Reloading the Flatness Correction Table
As a cautionary measure, the flatness corrections stored in the “Other” users file have been
provided on a matching serialized disk to be reloaded into the specific spectrum analyzer
in the event the flatness corrections are deleted from the instrument.
Insert the serialized disk into the instrument. Use the following keystrokes to reload the
serialized flatness corrections data from the disk to the specific instrument.
[FILE] > Load > Type > More 1 of 2 > Corrections > Dir Select > [ | ] >
Drive A > [ | ] > "H26CORR" > Load Now
Verification of the Flatness Correction Data
To verify the download, use the following keystrokes then verify that the data is
visible in the table.
[AMPLITUDE] > More 1 of 3 > Corrections > Other > Edit
Press [RETURN] so as not to disturb the data.
1. where CSET3 is the correction set value for "Other".