Specifications

Chapter 3 3-5
User Interface
Flatness Compensation Table
Flatness Compensation Table
The following sections will cover the following:
Accessing the Flatness Compensation Table” on page 3-5.
“GPIB Control of the Flatness Compensation Table” on page 3-5
“Reloading the Flatness Correction Table” on page 3-6.
“Recreating the Flatness Correction Table” on page 3-7.
Accessing the Flatness Compensation Table
During the Swept Flatness test routine at the factory, correction data points are collected
with the highband preamp in the OFF state. The test is then run again with the highband
preamp ON. The corrections are then applied to get the best flatness response.
The flatness compensation table is available for the user to apply corrections for an
improved flatness response while the highband preamp is on. The following keystrokes
will aid the user in accessing the flatness compensation table.
[AMPLITUDE] > More 1 of 3 > Corrections > Other > Correction > On | Off
NOTE
When sweeping from below 2.85 GHz to above 3.05 GHz, the corrections
table will cause amplitude errors in the 2.85 GHz – 3 GHz band. To
avoid these errors, set the start frequency above 2.85 GHz.
NOTE
The instrument does not automatically enable the Amplitude
Corrections, they must be enabled by the user where appropriate.
Unlike the option 1DS lowband preamp where the lowband preamp
flatness is corrected automatically, the Amplitude Correction “OTHER”
must be turned “ON” when the Preamp is turned “ON” and turned
“OFF” when the Preamp is turned “OFF.” Measurement errors will
result if this is not done.
GPIB Control of the Flatness Compensation Table
The following SCPI commands related to the Agilent E444xA Option 1DS and H26.
Refer to Agilent Technologies PSA Spectrum Analyzer User’s and Programmer’s Reference
Guide, to locate SCPI command subsystems and subsections that apply to the standard
functions of the Agilent Technologies E444xA PSA-Series Spectrum Analyzer.