Specifications

1-8 Chapter 1
Overview
Characteristics
-
Noise Figure Measurement Personality and Option H26
This section provides nominal performance information for the PSA series, Option 219,
Noise Figure Measurement as it relates to the Option H26, High Band Preamp.
CAUTION
Excessive Relay Switch wear can occur if the instrument is sweeping
from a start frequency of 3.05 GHz and below to above 3.05 GHz. This
is different than when operating in the SA Mode.
You need the following equipment to use this personality/hardware combination.
NOTE
The Noise Figure Measurement personality (Option 219) requires
Revision “b” or later of the Front End Driver assembly. This supplies
the +28 V output (labelled “NOISE SOURCE DRIVE OUT +28 V
(PULSED)” on the rear panel), which is needed to drive the noise
source. To see which version is installed on your PSA, press System,
Show Hdwr. If you have an earlier revision than Revision “b,” contact
your Agilent Technologies representative.
Refer to http://www.agilent.com/find/psa for further information.
Table 1-4 Hardware, Firmware and Software Requirements
Firmware Software Hardware
Revision
Number
Noise Figure
Measurement
Personality
Front End Driver
Board
Option 1DS
Internal
Pre-Amp
A.06.01 Option 219 Rev b or later Required
Table 1-5 Instrument Uncertainty for Noise Figure with Option H26
Description Specifications Supplemental Information
Instrument Uncertainty Nominally the same as for the
10 MHz to 3 GHz range
There are no
warranted
specifications
for Option H26.
Refer to Table
1-2 on page 1-4
for Nominal
Performance
H26 Preamp caution
a
Band Crossing caution
b
3 to 10 GHz
Well-controlled preselector
c
10 to 20 GHz
Good preselector stability
d
20 to 26.5 GHz
Preselector Drift Effects
e