Specifications
1-8 Chapter 1 
Overview
Characteristics
-
Noise Figure Measurement Personality and Option H26
This section provides nominal performance information for the PSA series, Option 219, 
Noise Figure Measurement as it relates to the Option H26, High Band Preamp.
CAUTION
Excessive Relay Switch wear can occur if the instrument is sweeping 
from a start frequency of 3.05 GHz and below to above 3.05 GHz. This 
is different than when operating in the SA Mode.
You need the following equipment to use this personality/hardware combination.
NOTE
The Noise Figure Measurement personality (Option 219) requires 
Revision “b” or later of the Front End Driver assembly. This supplies 
the +28 V output (labelled “NOISE SOURCE DRIVE OUT +28 V 
(PULSED)” on the rear panel), which is needed to drive the noise 
source. To see which version is installed on your PSA, press System, 
Show Hdwr. If you have an earlier revision than Revision “b,” contact 
your Agilent Technologies representative. 
Refer to http://www.agilent.com/find/psa for further information.
Table 1-4  Hardware, Firmware and Software Requirements
Firmware Software Hardware
Revision 
Number
Noise Figure 
Measurement 
Personality
Front End Driver 
Board
Option 1DS 
Internal 
Pre-Amp
≥ A.06.01 Option 219 Rev b or later Required
Table 1-5  Instrument Uncertainty for Noise Figure with Option H26
Description Specifications Supplemental Information
Instrument Uncertainty Nominally the same as for the 
10 MHz to 3 GHz range
There are no 
warranted 
specifications 
for Option H26. 
Refer to Table 
1-2 on page 1-4 
for Nominal 
Performance
H26 Preamp caution
a
Band Crossing caution
b
3 to 10 GHz
Well-controlled preselector
c
10 to 20 GHz
Good preselector stability
d
20 to 26.5 GHz
Preselector Drift Effects
e