Specifications

16
Account for mixer characteristics. See Hint # 6.
Measure the same sideband(s) that will be used in
the application.
For double-sideband measurements, select a LO
frequency close to the RF band of interest.
For single-sideband measurements, select a LO far
from the RF band of interest, if possible.
Choose the LO to suit the mixer.
Filter the LO if necessary to diminish spurious signals
and broadband noise.
Keep the LO outside of the measurement bandwidth
if possible.
Filter the IF if necessary to remove LO harmonics
created inside the mixer.
Filter the RF to prevent unwanted mixing.
Test for DSB error by changing IF.
Experiment with different LOs to get the most
accurate (i.e. lowest) noise figure.
Document a frequency plan to evaluate which of the
above precautions are necessary.
Enter proper measurement correction.
See Hint # 7.
Calibrate regularly.
Keep overall gain below the instrument’s spec.
Filter out-of-band power.
If 2nd stage effect is large, add a low-noise pre-amp
(with proper connectors).
Choose the optimal bandwidth. See Hint # 8.
Select a measurement bandwidth no larger than the
pass band of the DUT.
Account for path losses. See Hint # 9.
Avoid adapters as much as possible.
If used, enter their losses into the instrument.
Account for the temperature of measurement
components. See Hint # 10.
Enter physical temperatures of the noise source and
the components of the measurement setup into the
instrument.
Appendix A:
Checklist
Use this checklist to assist with locating the hint relating to
specific issues or considerations. Reading the 10 Hints
Application Note sequentially is not necessary.
Select the appropriate noise source. See Hint # 1.
Use a low ENR source whenever possible.
Avoid adapters between the noise source and DUT.
Double check manually entered ENR values.
Reduce EMI influence. See Hint # 2.
Use clean, undamaged connectors.
Use threaded connectors.
Use double shielded cables.
Use shielded GPIB cables.
Use a screened room.
Use shielding.
Use an analyzer with minimal electromagnetic
emissions.
Minimize mismatch uncertainties. See Hint # 3.
Use an attenuator or isolator if needed.
Use a pre-amp if needed.
Use averaging to minimize display jitter.
See Hint # 4.
Select enough averages to stabilize the measurement.
Use “trace averaging” first to spot measurement setup
problems soonest.
Look for spikes or even small steps in the display that
indicate RF interference.
If time constrained, use more averaging during
calibration than during DUT measurement.
Avoid non-linearities. See Hint # 5.
Avoid the following:
Circuits with phase lock loops
Circuits that oscillate
Amplifiers or mixers operating near saturation
AGC circuits or limiters
High-gain DUTs without in-line attenuation
Power supply drifts
DUTs or instruments that have not warmed up
Logarithmic amplifiers