User`s guide

Appendix B GPIB Commands 307
TRIGGER (GET)
TRIGGER (GET)
If triggering is armed (see TARM command), the TRIGGER command (Group
Execute Trigger) triggers the multimeter once, and then holds triggering.
Syntax TRIGGER 7
TRIGGER 722
Remarks The TRIGGER command generates a single trigger just as if the TRIG SGL
command was executed. It will not, however, trigger the multimeter if triggering
is not armed (TARM command).
If subprogram memory execution is suspended by the PAUSE command
(multimeter command set), the TRIGGER command resumes subprogram
execution but does not generate a single trigger.
Examples TRIGGER 7 !SENDS GROUP EXECUTE TRIGGER (GET)
TRIGGER 722 !SENDS GROUP EXECUTE TRIGGER (GET) TO THE DEVICE AT ADDRESS 22