User`s guide

BEEP
164 Chapter 6 Command Reference
BEEP
Controls the multimeter's beeper. When enabled, the beeper emits a 1 kHz beep
if an error occurs.
Syntax BEEP [control]
control The control parameter choices are:
Power-on control = last programmed value.
Default control = ONCE.
Remarks The multimeter stores the control parameter in continuous memory (the
parameter is not lost when power is removed).
Query Command. The BEEP? query command returns the present beeper
mode. Refer to "Query Commands" near the front of this chapter for more
information.
Related Commands: TONE
Example OUTPUT 722;"BEEP OFF" !DISABLES THE BEEPER
CAL
This is a calibration command. Refer to the 3458 Calibration Manual for details.
CALL
Call Subprogram. Executes a previously stored subprogram.
Syntax CALL [name]
name Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the SUB command for details.
Power-on name = none.
control
Parameter
Numeric
Query
Equiv. Description
OFF 0 Disables the beeper
ON 1 Enables the beeper
ONCE 2 Beeps once, then returns to previous mode (either OFF
or ON)