CALNUM?
Description Calibration Number Query. Returns a decimal number indicating the number
of times the multimeter has been adjusted.
Syntax CALNUM?
Remarks • The calibration number is incremented by 1 whenever the multimeter is
unsecured and adjusted. If autocal is secured, the calibration number is
also incremented by 1 whenever an autocal is performed; if unsecured,
autocal does not affect the calibration number.
• The calibration number is stored in cal-protected memory and is not lost
when power is removed.
• The multimeter was adjusted before it left the factory which increments
the CALNUM. When you receive the multimeter, read the calibration
number to determine its initial value.
• Related Commands: CAL. CALSTR. SCAL
Example 10 OUTPUT 722;"CALNUM?" !READS CALIBRATION NUMBER
20 ENTER 722;A !ENTERS RESPONSE INTO COMPUTER
30 PRINT A !PRINTS RESPONSE
40 END