CALNUM?
Description Calibration Number Query. Returns a decimal number indicating the number
of times the multimeter has been adjusted.
Syntax CALNUM?
Remarks The calibration number is incremented by 1 whenever the multimeter is
unsecured and adjusted. If autocal is secured, the calibration number is
also incremented by 1 whenever an autocal is performed; if unsecured,
autocal does not affect the calibration number.
The calibration number is stored in cal-protected memory and is not lost
when power is removed.
The multimeter was adjusted before it left the factory which increments
the CALNUM. When you receive the multimeter, read the calibration
number to determine its initial value.
Related Commands: CAL. CALSTR. SCAL
Example 10 OUTPUT 722;"CALNUM?" !READS CALIBRATION NUMBER
20 ENTER 722;A !ENTERS RESPONSE INTO COMPUTER
30 PRINT A !PRINTS RESPONSE
40 END