Programming instructions
57
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Test Performed By
Keysight Model 3458A Digital Multimeter
Serial Number
Date
OHMS TESTS
CAL? 60
TEMP?
Difference - (must be less than 5 degrees C)
Perform an ACAL OHMS
Test # 3458A
Input
3458A
Range
Transfer Standard
Reading
Unit Under Test
Reading
Difference Limit (Std) Limit (Opt 002) Pass Fail
2-Wire Function Offset Test
1Short
10 Ω
N/A N/A
00.25007
Ω 00.25007 Ω
4-Wire Function Offset
Test (Rear Terminals)
1
Short 10 Ω
N/A N/A
00.00007
Ω 00.00007 Ω
4-Wire Function Gain Test
1
10 Ω 10 Ω 00.00028 Ω 00.00028 Ω
2
100 Ω 100 Ω 000.00231 Ω 000.00231 Ω
3
1 KΩ 1 KΩ 0.0000142 KΩ 0.0000142 KΩ
4
10 KΩ 10 KΩ 00.000142 KΩ 00.000142 KΩ
5
100 KΩ 100 KΩ 000.00142 KΩ 000.00142 KΩ
6
1 MΩ 1 MΩ 0.0000209 MΩ 0.0000209 MΩ
7
10 MΩ 10 MΩ 00.000703 MΩ 00.000703 MΩ
7
a
10 MΩ 10 MΩ 00.000703 MΩ 00.000703 MΩ