Programming instructions

53
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Test Performed By
Keysight Model 3458A Digital Multimeter
Serial Number
Date
AC VOLTAGE TESTS
Perform an ACAL AC
Test # 3458A
Input
3458A
Range
Transfer Standard
Reading
Unit Under Test
Reading
Difference Limit (Std) Limit (Opt 002) Pass Fail
1 100 mV, 1KHz 100 mV 000.0250 mV 000.0250 mV
2
1 V, 1KHz 1 V 0.000250 V 0.000250 V
3 1 V, 1KHz 10 V 00.00096 V 00.00096 V
4
10 V, 20 Hz 10 V 00.01338 V 00.01338 V
5
10 V, 1KHz 10 V 00.00250 V 00.00250 V
6
10 V, 20KHZ 10 V 00.00272 V 00.00272 V
7
10 V, 100 KHz 10 V 00.05372 V 00.05372 V
8
10 V, 1MHz 10 V 00.55450 V 00.55450 V
9
100 V, 1KHz 100 V 000.0364 V 000.0364 V
10
700 V, 1 KHz 1000 V 0000.544 V 0000.544 V