Programming instructions

51
PERFORMANCE TEST CARD - 1 YEAR LIMITS
Test Performed By
Keysight Model 3458A Digital
Multimeter Serial Number Date
DC VOLTAGE TESTS
CAL? 59
TEMP?
Difference - (must be less than 5 degrees C)
Perform an ACAL DVC
Test # 3458A
Input
3458A
Range
Transfer Standard
Reading
Unit Under Test
Reading
Difference Limit (Std) Limit (Opt 002) Pass Fail
OFFSET TESTS
(NOTE: Math Null is Disabled)
1Short
100 mV
N/A N/A 000.00106 mV 000.00106 mV
2 Short 1 V N/A N/A 0.00000106 V 0.00000106 V
3Short
10 V
N/A N/A 00.0000023 V 00.0000023 V
4 Short 100 V N/A N/A 000.000036 V 000.000036 V
5
Short 1000 V
N/A N/A 0000.00010 V 0000.00010 V
GAIN TESTS
1 100 mV 100 mV 000.00212 mV 000.00188 mV
2 1 V 1 V 0.00000998 V 0.00000740 V
3 1 V 10 V 00.0000111 V 00.0000085 V
4 -1 V 10 V 00.0000111 V 00.0000085 V
5 -10 V 10 V 00.0000892 V 00.0000624 V
6 10 V 10 V 00.0000892 V 00.0000624 V
7 100 V 100 V 000.001114 V 000.000853 V
8 1000 V 1000 V 0000.02396 V 0000.01934 V