Programming instructions

Chapter 4 Performance Verification Tests 41
16. After all DC gain tests have been performed, calculate and record in
the column provided. The difference (absolute value) between the
transfer standard DMM reading and the unit under test reading for
each of the test points.
17. If any of the differences calculated are beyond the specified limits,
refer to Chapter 3. "Adjustment Procedures", to make adjustments.
Table 2. DCV Gain Performance Tests
DC Gain Test
Number
DMM Range Source Output
1 100 mV 100 mV
2 1 V 1 V
3 10 V 1 V
4 10 V -1 V
5 10 V -10 V
6 10 V 10 V
7 100 V 100 V
8
a
a.NOTE: After completing test 8, decrease the 1000V verification
source output to 0V before disconnecting.
1000 V 1000 V