User`s guide
58 34450A User’s Guide
2 Functions and Features
Reset/Power-On State
The table below summarizes the 34450A settings as received
from the factory, following power cycling, and following the
*RST command received over the USB remote interface.
Non- volatile, user customizable behavioral differences are
shown in BOLD.
Tab le 2-7 Reset/Power-on state
Parameter Factory Setting Power-On / Reset State
Measurement Configuration
Function DCV DCV
Range AUTO AUTO
Resolution 5½ digits 5½ digits
Temp e rat u r e U n i ts °C User setting
Math Operations
Math State, Function Off, Null Off, Null
Math Registers Cleared Cleared
dBm Reference Resistance 600 Ω User setting
Trigger Operation
Trigger Source
[1]
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
Auto Trigger (Local Mode)
IMMediate (Remote Mode)
System-Related Operation
Power-Down Recall Disabled User Setting
Stored States 0-5 cleared No Change
Beeper On User Setting
Display On On
Remote/ Local State Local Local
Keyboard
[1]
Unlocked, Local key enabled Unlocked, Local key enabled