User`s guide

48 34450A User’s Guide
2 Functions and Features
Tab le 2-5 Utility menu available settings
Feature Default Available Settings Description Remote Command
BUZZER ON ON or OFF
Enables or disables Diode, Stats, Limit,
and Hold beep operations.
Turning off the buzzer does not disable
the front panel keys beep operation and
Continuity beep operation.
Refer to “The beeper” on page 55 for
more info.
SYSTem:BEEPer:STATe
I/O USB
USB, GPIB, or
RS232
Disables or enables the GPIB, USB,
or RS232 remote interface
If GPIB is selected, refer to “GPIB
Utility Sub-Menu” on page 53
If RS232 is selected, refer to “RS232
utility sub-menu” on page 51
When all I/O are disabled, DISABLE
is shown on the setting.
SYSTem:COMMunicate:
ENABle <mode>,
<interface>
TEMP UNIT °C °C or °F
Selects the unit for temperature
measurements
UNIT:TEMPerature
<units>
LANGUAGE L1 L1 or L2
L1 represents Keysight Mode
L2 represents Fluke 45/8808A Mode.
Refer to “Enabling the code
compatibility function” on page 73 for
more info.
SYSTem:LANGuage
INPUT Z 10M 10M or HIGH Z
Sets the input impedance for DCV
measurements (HIGH Z is selectable
for 100 mV and 1 V ranges only)
[SENSe:]VOLTage[:DC
]:IMPedance:AUTO
<mode>
SELF TEST OFF ON or OFF
ON enables an immediate self-test of
the multimeter. Returns to normal
operation after completing the
self-test.
*TST?
P-ON RESET ON ON or OFF
Disables or enables the automatic
recall of the powered down state when
power is turned on
MEMory:STATe:RECall
:AUTO
OCOMP OFF ON of OFF
Enables or disables the offset
compensation for resistance
measurement
[SENSe:]RESistance:
OCOMpe
nsated <mode>