User`s guide

108 34450A User’s Guide
4 Specifications
Supplemental measurement specifications
Tab le 4-7 Supplemental measurement specifications
DC Voltage
Measuring Method:
Sigma Delta A-to-D converter
Input Resistance:
>10 GΩ ± 2% range (Selectable 100 mV, 1 V ranges)
10 MΩ ± 2% range (typical)
Input Protection:
1000 V on all ranges (HI terminal)
Resistance
Measurement Method:
2-wire or 4-wire Ohms
Open-Circuit Voltage
Limited to < 2.8 V
Input Protection:
1000 V on all ranges (HI terminal)
DC Current
Shunt Resistance:
1 Ω for 10 mA, 100 mA
90 Ω for 100 μA, 1 mA
0.01 Ω for 1 A, 10 A
Input Protection:
Externally accessible at Rear Panel 0.4 A, 500 V fuse for I terminal
Internal 11 A, 1000 V fuse for 10 A terminal