User`s guide
102 34450A User’s Guide
4 Specifications
Continuity
[3]
1000 Ω 0.5 mA - 0.05 + 0.03 0.005 + 0.005
Diode Test
[4]
1.0000 V 0.5 mA - 0.05 + 0.03 0.005 + 0.005
[1] 20% over range on all ranges except 1000 VDC and 10 A range
[2] Specifications are for 4-wire Ω or 2-wire Ω using NULL function. If without NULL function, add 0.2 Ω additional error.
[3] Continuity thresholds is fixed at less than 10 Ω
[4] Specifications are for the voltage measured at the input terminals only
Tab le 4-1 DC accuracy ± (% of reading + % of range) (continued)
Function Range
[1]
Test Current or
Burden Voltage
Input Impedance
1 Year
23 °C ± 5° C
Temperature
Coefficient
0 °C – 18 °C
28 °C – 55 °C