Technical data
1 Performance Tests and Calibration
Calibration Message
40 34450A Service Guide
Calibration Message
The instrument allows you to store a message in calibration memory. For
example, you can store such information as the date when the last
calibration was performed, the date when the next calibration is due, the
instrument's serial number, or even the name and phone number of the
person to contact for a new calibration. The calibration message may
contain up to 40 characters.
You can record a calibration message only when the instrument is
unsecured. You can read the calibration message whether the instrument
is secured or unsecured.
To store a calibration message use the
CALibration:STRing command
and to read the calibration message, use the
CALibration:STRing?
command from the remote interface.
To read the calibration count
You can query the instrument to determine how many calibrations have
been performed. Note that your instrument was calibrated before it left
the factory. When you receive your instrument, read the count to
determine its initial value.
The count value increments by one for each calibration point, and a
complete calibration will increase the value by many counts. The
calibration count increments up to a maximum of 2
32
- 1 after which it
rolls over to 0. The calibration count can be read either remotely or from
the front panel. To read the calibration count from remote, use the
CALibration:COUNt? command. Use the following procedure to read
the calibration count from the front panel.
1 In the Utility menu/Calibration mode display will show the calibration
count.