Technical data
Performance Tests and Calibration 1
Calibration Process
34450A Service Guide 27
Using the front panel for adjustments
This section describes the process used to perform adjustments from the
front panel. Refer to the 34450A Programmer's Reference online help for
remote interface commands.
Selecting the Adjustment Mode
1 Press > to enter the Utility menu.
2 Select CALIBRATION and press to enter Calibration menu.
3 Select CAL MODE and press to enter Calibration mode. Note
that CAL MODE will not exist in the menu if the instrument is
secured.
Entering adjustment values
In the multimeter adjustment procedures, all calibration point has been
preset. You only need to select the calibration range to start the
calibration process.
Press or to select calibration range.
Press to execute calibration.
Storing the calibration constants
The calibration constants are automatically stored at the end of the
calibration process for each range.
Aborting a calibration in progress
Sometimes it may be necessary to abort a calibration after the procedure
has already been initiated. You can abort a calibration at any time by
pressing