Technical data

1 Performance Tests and Calibration
Performance Verification Tests
18 34450A Service Guide
1 V
1 kHz
1 V Q ±3 mV
1 V
10 kHz
1 V ±3 mV
1 V
30 kHz
1 V ±18 mV
1 V
100 kHz
1 V ±33 mV
10 V
20 Hz
10 V ±0.11 V
10 V
45 Hz
10 V ±30 mV
10 V
1 kHz
10 V ±30 mV
10 V
10 kHz
10 V ±30 mV
10 V
30 kHz
10 V Q ±0.18 V
10 V
100 kHz
10 V ±0.33 V
100 V
45 Hz
100 V ±0.3 V
100 V
1 kHz
100 V ±0.3 V
100 V
10 kHz
100 V Q ±0.3 V
100 V
30 kHz
100 V ±1.8 V
100 V
100 kHz
100 V ±3.3 V
750 V
45 Hz
750 V ±2.25 V
750 V
1 kHz
750 V ±2.25 V
750 V
10 kHz
750 V ±2.25 V
[1] Select Slow Mode 5½ digit measurement resolution
Q= Quick performance verification test points
Table 1- 7 AC volts verification test
Vrms Input Frequency Function
[1]
Range Quick Check
Error from
Nominal 1 Year