Technical data
Performance Tests and Calibration 1
Performance Verification Tests
34450A Service Guide 17
AC voltage verification test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Connect the calibrator to the front panel HI and LO input terminals
as shown in the figure below:
2 Select the AC voltage function.
3 Select each range in the order shown below. Provide the indicated
input voltage and frequency. Compare measurement results to the
appropriate test limits shown in the table. (Be certain to allow for
appropriate source settling.)
Table 1- 7 AC volts verification test
Vrms Input Frequency Function
[1]
Range Quick Check
Error from
Nominal 1 Year
100 mV 20 Hz AC voltage 100 mV ±1.1 mV
100 mV
45 Hz
100 mV ±0.3 mV
100 mV 1 kHz 100 mV ±0.3 mV
100 mV
10 kHz
100 mV ±0.3 mV
100 mV
30 kHz
100 mV ±1.8 mV
100 mV
100 kHz
100 mV ±3.3 mV
1 V
20 kHz
1 V ±11 mV
1 V
45 kHz
1 V ±3 mV