Technical data

1 Performance Tests and Calibration
Test Considerations
6 34450A Service Guide
Input connections
Test connections to the instrument are best accomplished using the dual
banana plug with copper wire shorted between two terminals for
low- thermal offset measurement. Shielded, twisted- pair, PTFE
interconnect cables of minimum length are recommended between the
calibrator and the multimeter. Cable shields should be earth ground
referenced. This configuration is recommended for optimal noises and
settling time performance during calibration.