Specifications
To Use Reading Hold
The reading hold feature allows you to capture and hold a stable reading
on the display. When a stable reading is detected, the meter emits a beep
and holds the value on the display.
• Reading hold has an adjustable sensitivity band to allow you to select
which readings are considered stable enough to be displayed. The
band is expressed as a percent of reading on the selected range. The
meter will capture and display a new value only after three
consecutive readings are within the band.
• The default band is 0.10% of reading. After enabling reading hold,
you can choose a different band by pressing
Shift > (Menu Recall).
This takes you to the “READ HOLD” command in the TRIG MENU
(only if reading hold is enabled).
Go down to the “parameter” level, and then select a value: 0.01%,
0.10%, 1.00%, or 10.00% of reading.
• The sensitivity band is stored in volatile memory; the meter sets the
band to 0.10% of reading when power has been off or after a remote
interface reset.
Toggles between autotrigger
and reading hold
* (sample) annunciator is on
during each measurement
Chapter 2 Front-Panel Operation
To Use Reading Hold
48