Specifications
511 RS-232 framing error
512 RS-232 overrun error
513 RS-232 parity error
514 Command allowed only with RS-232
There are three commands which are only allowed with the
RS-232
interface: SYSTem:LOCal, SYSTem:REMote, and SYSTem:RWLock.
521 Input buffer overflow
522 Output buffer overflow
531 Insufficient memory
There is not enough memory to store the requested number of readings in
internal memory using the INITiate command. The product of the
sample count (SAMPle:COUNt) and the trigger count (TRIGger:COUNt)
must not exceed 1024 readings.
532 Cannot achieve requested resolution
The meter cannot achieve the requested measurement resolution. You
may have specified an invalid resolution in the CONFigure or MEASure
command.
540 Cannot use overload as math reference
The meter cannot store an overload reading (9.90000000E+37) as the
math reference for null measurements. The math state is turned off as a
result of this condition.
550 Command not allowed in local
The meter received a READ? command while in the local mode for
RS-232
operation. You should always execute the SYSTem:REMote command
before sending other commands over the interface.
552 Command not allowed in current function
Chapter 5 Error Messages
Execution Errors
206