Specifications
[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe:AUTO {OFF | ON}
[SENSe:]
FRESistance | RESistance
:RANGe:AUTO {OFF | ON}
This command disables or enables autoranging for the function.
Autorange thresholds: Down range at <10% of range; Up range at >120%
of range. [volatile memory]
[{SENSe1: | SENSe2:}]VOLTage[:DC]:RANGe:AUTO?
[SENSe:]
FRESistance | RESistance
:RANGe:AUTO?
This command queries the autorange setting for the specified function.
Returns “0” (
OFF) or “1” (ON).
[SENSe1: | SENSe2:]VOLTage:[DC:]RESolution {<
resolution
> | MIN | MAX}
[SENSe:]
FRESistance | RESistance
:RESolution {<
resolution
> | MIN | MAX}
This command sets the resolution for the specified function. Specify the
resolution in the same units as the measurement function, not in number
of digits. MIN selects the smallest value accepted, which gives the most
resolution. MAX selects the largest value accepted which gives the least
resolution. [volatile memory]
[SENSe1: | SENSe2:]VOLTage:[DC:]RESolution? [{MIN | MAX}]
[SENSe:]
FRESistance | RESistance
:RESolution? [{MIN | MAX}]
This command queries the resolution for the selected function.
Chapter 4 Remote Interface Reference
Setting the Function, Range, and Resolution
138