User`s guide

86 34410A/11A/L4411A User’s Guide
2Features and Functions
Thousands Separator (34410A/11A) On
Display State (34410A/11A) On
Reading Memory Cleared
Error Queue Cleared at Power
On.
Not cleared
by a front panel or
remote interface Reset command
Stored States No change
Calibration State Secured
Input/Output Configuration Factory Setting
Enable Interfaces:
1
LAN Enabled
GPIB Enabled
USB Enabled
Lan Settings:
2
DHCP On
Auto IP On
IP Address 169.254.4.10
Subnet Mask 255.255.0.0
Default Gateway 0.0.0.0
DNS Server 0.0.0.0
Host Name A-34410A-nnnnn (for 34410A)
3
A-34411A-nnnnn (for 34411A)
A-L4411A-nnnnn (for L4411A)
LAN Services
1
Enable All
GPIB Settings:
GPIB Address 22
1 Interface enable or LAN service changes take effect only after you cycle power.
2 LAN setting changes require a LAN restart. From SCPI you must cycle power.
3 Where nnnnn represents the last 5 digits of the instrument serial number.
SystemRelated Operations Factory Setting
NOTE
The power-on/reset state may be different than that in the above tables if you have
enabled the power-on state recall mode (PWR-ON AUTO set to On from the Utility menu).
Also, you can recall one of four stored instrument states other than the power-on state. See
“Multimeter State Storage” on page 48 for further information.
UG_ED5.book Page 86 Thursday, March 1, 2012 11:28 AM