82 34410A/11A/L4411A User’s Guide
2Features and Functions
System-Related Operations
The utility menu accesses the following system–related operations: self–test,
display of error conditions, calibration, and instrument information. These
features are described in the subsections that follow.
The utility menu is also used to access the following three menus:
REMOTE I/O menu (see Chapter 3, “Remote Interface Configuration” for a
complete description of the remote interface features).
STORE/RECALL menu (see “Multimeter State Storage” on page 48 for a
complete description of the state storage feature).
MISC SETTINGS menu (see Miscellaneous Configuration Settingson
page 60 for a complete description of these features).
SelfTe s t
A poweron selftest occurs automatically when you turn on the instrument.
This limited test assures you that the multimeter is operational. This self–test
does not perform the extensive self test described below.
A complete self–test performs a series of internal tests, and takes about five
seconds to execute.
If the power–on or complete self–test fails, an error is stored in the error
queue. See the Agilent 34410A/11A/L4411A Service Guide for more
information on obtaining service.
Following the complete self–test, the instrument issues a Factory Reset
(*RST) command.
Front Panel Operation: Press
UTILITY MENU > SELF-TEST
If the test fails, the Error annunciator will light.
Remote Interface Operation: The following command returns “+0“ if the
self–test is successful or “+1 if it fails.
*TST?
Refer to the Agilent 34410A/11A/L4411A Programmer’s Reference Help for
a complete description and syntax for this command.