User`s guide

34410A/11A/L4411A User’s Guide 77
Features and Functions 2
Trigger Slope
You may select whether the multimeter uses the rising edge (POS) or falling
edge (NEG) of the external trigger signal to trigger a reading, or
(independently) for the voltmeter complete output signal. The default for both
is NEG.
Front Panel Operation: Press
TRIGGER > SETUP > N SAMPLES > TRIG DELAY > TRIG SLOPE > VMC SLOPE
For each of TRIG SLOPE and VMC SLOPE, you can select either NEG or POS:
TRIG SLOPE sets the slope of the Ext Trig input signal (NEG or POS).
VMC SLOPE sets the slope of the VM Comp output signal (NEG or POS).
Then step through or exit the configuration menu.
Remote Interface Operation: The following command selects the trigger slope
for the external trigger (Ext Trig) connector:
TRIGger:SLOPe {POSitive>NEGative}
The following query returns the selected trigger slope for the Ext Trig
connector:
TRIGger:SLOPe?
This query returns POS or NEG.
The following command selects the signal slope for the voltmeter complete
(VM Comp) signal:
OUTPut:TRIGger:SLOPe {POSitive>NEGative}
The following query returns the selected trigger slope for the VM Comp
connector:
OUTPut:TRIGger:SLOPe?
This query returns POS or NEG.
Refer to the Agilent 34410A/11A/L4411A Programmer’s Reference Help for
a complete description and syntax for this command.
UG_ED5.book Page 77 Thursday, March 1, 2012 11:28 AM