User`s guide

34410A/11A/L4411A User’s Guide 111
Measurement Tutorial 4
The following table shows the typical error for various pulse waveforms as a
function of input pulse frequency:
This table gives an additional error for each waveform, to be added to the
value from the accuracy table provided in Chapter 5, “Specfications.”
The 34410A/11A/L4411A specifications are valid for CF
10, provided there
is insignificant signal energy above the 300 kHz bandwidth for voltage, or
the 10 kHz bandwidth for current. Multimeter performance is not specified
for CF > 10, or when significant out-of-band signal content is present.
Example:
A pulse train with level 1 V
rms
, is measured on the 1 V range. It has pulse
heights of 3 V (that is, a Crest Factor of 3) and duration 111 μs. The prf can be
calculated to be 1000 Hz, as follows:
Thus, from the table above, this ac waveform can be measured with 0.18
percent additional error.
Typical error for square wave, triangular wave, and pulse trains of CF=3, 5, or 10
prf square wave triangle wave CF=3 CF=5 CF=10
200
0.02% 0.00% 0.04% 0.09% 0.34%
1000
0.07% 0.00% 0.18% 0.44% 1.71%
2000
0.14% 0.00% 0.34% 0.88% 3.52%
5000
0.34% 0.00% 0.84% 2.29% 8.34%
10000
0.68% 0.00% 1.75% 4.94% 26.00%
20000
1.28% 0.00% 3.07% 8.20% 45.70%
50000
3.41% 0.04% 6.75% 32.0% 65.30%
100000
5.10% 0.12% 21.8% 50.6% 75.40%