User`s guide
Examples
The following command configures the instrument for 2-wire resistance
measurements, triggers the meter to take a reading, and then transfers the
reading to the instrument output buffer. The default range (autorange) and
resolution (1 PLC) are used for the measurement.
MEAS:RES?
Typical Response: +1.32130000E+04
The following command configures the instrument for 2-wire resistance
measurements, triggers the meter to take a reading, and then transfers the
reading to the instrument output buffer. The 1 kΩ range is selected with 0.1
Ω resolution.
MEAS:RES? 1000,0.1
Typical Response: +4.27150000E+02