Remarks
The range and resolution are fixed for continuity tests. The range is 1 kΩ
(a 2-wire resistance measurement).
The meter beeps (even if beep is disabled) for each measurement that is
less than or equal to the continuity threshold (≤10 Ω), and the actual
resistance reading is displayed on the front panel.
From 10 Ω to 1.2 kΩ the meter displays the actual resistance reading with
no beep.
If the reading exceeds 1.2 kΩ, the meter displays "OPEN" on the front
panel (no beep).
The MEASure:CONTinuity? command returns the measured resistance,
regardless of its value.
Either the *RST command or the Instrument Preset (the SYSTem:PRESet
command) will set all measurement parameters to their factory settings,
clear reading memory, and clear all stored statistical data.