The instrument has five storage locations in non-volatile memory to store
instrument states. The instrument uses location 0 to automatically hold
the state of the instrument at power down. You can store the instrument
state in location 1, 2, 3, or 4, but you can only recall a state from a
location that contains a previously stored state. You can use location 0 to
store a fifth instrument state, but keep in mind that location 0 is
automatically overwritten when power is cycled.
When shipped from the factory, storage locations 1 through 4 are empty
and the instrument is configured to automatically recall the power-down
state (location 0) when power is restored. You can change the factory
configuration such that a Factory Reset (*RST command) is issued when
power is restored.
You can assign a user-defined name to each of locations 0 through 4. See
the MEMory:NAME command for details.
A Factory Reset (*RST command) does not affect the configurations stored
in memory. Once a state is stored, it remains until it is overwritten or
specifically deleted.