The following table lists the bit definitions for the Standard Event Register.
Bit Number
Decimal
Value
Definition
0 Operation
Complete
1
All commands prior to and
including *OPC have been
executed.
1 Not Used Not Used "0" is returned.
2 Query Error 4
The instrument tried to read
the output buffer but it was
empty. Or, a new command line
was received before a previous
query has been read. Or, both
the input and output buffers
are full.
3 Device-
Specific Error
8
A device-specific error,
including a self-test error or
calibration error, occurred (an
error in the -300 range or any
positive error has been
generated). For a complete
listing of the error messages,
see SCPI Error Messages.
4 Execution
Error
16
An execution error occurred (an
error in the -200 range has
been generated).
5 Command
Error
32
A command syntax error
occurred (an error in the -100
range has been generated).
6 Not Used Not Used "0" is returned.
7 Power On 128
Power has been turned off and
on since the last time the event
register was read or cleared.