To determine the NPLCs for a specified range and resolution, locate the range
in the left column. Measurement units do not apply (i.e. for 1V or 1A, use the
1 range; for 100V or 100 ohms, use the 100 range). From the range value,
follow the corresponding row of resolutions to the right until the specified
resolution lies within two adjacent columns.
The NPLC value at the top of the left-adjacent column is the NPLC "setting" for
the given range and resolution.
For example: for a specified 10 VDC range and 10uV resolution, locate range
10 (1E1) in the table. Moving to the right, the 10uV resolution (10E-6) falls
between the adjacent columns containing 7E-6 and 15E-6. The corresponding
NPLC value is found by following the left-adjacent column (the 7E-6 column) to
the top where the NPLC setting is 0.2.