User`s guide
autoranging. The range in use cannot be predicted, nor can the
delay. For temperature measurements, use TRIGger:DELay? to
determine the current delay.
Resistance (4-wire):
Range
Trigger Delay
≤0.02
PLC
0.06
PLC or
0.2
PLC
≥1
PLC
100 Ω 1.0 ms 1.0 ms 1.5 ms
1 kΩ 1.0 ms 1.0 ms 1.5 ms
10 kΩ 1.0 ms 1.0 ms 1.5 ms
100 kΩ 1.0 ms 1.0 ms 1.5 ms
1 MΩ 10 ms 10 ms 15 ms
10 MΩ 100 ms 100 ms 100 ms
100 MΩ 100 ms 100 ms 100 ms
1 GΩ 100 ms 100 ms 100 ms
Four-wire temperature measurements actually use the 4-wire
resistance function. The above table applies to 4-wire temperature