User`s guide

Automatic Trigger Delays
If you do not specify a trigger delay in the TRIGger:DELay command, the
instrument selects a delay for you. The delay is determined by the function,
range, integration time, and ac filter setting as shown below. See the
TRIGger:DELay and TRIGger:DELay:AUTO commands.
For measurements from the remote interface, the automatic trigger
delay values are as shown in the following tables. These values are
also used for single-trigger, reading-hold, and level-trigger
measurements from the front panel. However, for auto-trigger
measurements from the front panel only, the automatic trigger
delay is 50 ms for all measurement functions.
DC Voltage (for all ranges):
Integration
Time
Trigger
Delay
1 PLC 160 µs
0.06 PLC or
0.2 PLC
130 µs
0.02 PLC 100 µs
DC Current (for all ranges):
Integration
Time
Trigger
Delay
1 PLC 1.5 ms
0.06 PLC or
0.2 PLC
1.0 ms