User`s guide

Remarks
The condition register bits reflect the current condition. If a condition goes
away, the corresponding bit is cleared in the condition register.
The following table lists the bit definitions for the Standard Operation
Register.
Bit Number
Decimal
Value
Definition
0
Calibration in
Progress
1
Instrument is performing a
calibration.
1-3 Not Used Not Used "0" is returned.
4 Measuring 16
Instrument is initiated, and is
making, or about to make a
measurement.
5
Waiting for
Trigger
32
Instrument is waiting for a
trigger.
6-7 Not Used Not Used "0" is returned.
8
Configuration
Change
256
Instrument configuration has
been changed, either from the
front panel or from SCPI.
9
Memory
Threshold
512
Programmed number of readings
have been stored in reading
memory (see
DATA:POINts:EVENt:THReashold
command).
10
Instrument
Locked
1024
If a remote interface (GPIB,
USB, or LAN) has a lock (see
SYSTem:LOCK:REQuest?
command), this bit will be set.
When a remote interface
releases the lock (see
SYSTem:LOCK:RELease
command), this bit will be
cleared.