User`s guide

Remarks
Affects both 2-wire and 4-wire RTD measurements.
Only Type 85 RTDs are suppported for RTD temperature measurements.
The instrument supports RTDs with Alpha = 0.00385 (DIN/IEC 751) using
ITS-90 software conversions.
"PT100" is a special label that is sometimes used to refer to an RTD with
Alpha = 0.00385 and Ro = 100Ω.
The instrument sets the RTD type to "85" after a Factory Reset (*RST
command) or an Instrument Preset (SYSTem:PRESet command).