Technical data
Performance Tests and Calibration 4
34405A User’s and Service Guide 79
Optional AC Voltage Performance Verification Test
Configuration: AC Volts (CONFigure[:VOLTage]:AC)
1 Select the AC Voltage function.
2 Select each range in the order shown below. Provide the indicated input
voltage and frequency. Compare measurement results to the appropriate
test limits shown in the table. (Be certain to allow for appropriate
source settling.)
[1] Select 5½- digit measurement resolution
Table 11 Optional AC Voltage Performance Verification Test
Vrms Input Frequency Function
[1]
Range
Error from Nominal
1 year
1V 45Hz AC Voltage 1V ±11mV
1V 1kHz 1V ±3mV
1V 10kHz 1V ±3mV
1V 30kHz 1V ±11mV
1V 100kHz 1V ±32mV
10V 1kHz 10V ±30mV
1V 1kHz 10V ±12mV
0.1V 1kHz 10V ±10.2mV